Monday, 3 February 2020

Diffusion Module


Process and visualize diffusion images, with the diffusion weighted imaging (DWI) model, the diffusion tensor imaging (DTI) model, and the new constrained spherical deconvolution (CSD) model, addressing the issue of crossing fibers.

constrained spherical deconvolution (CSD)

The corticospinal tract (part of the motor network) and arcuate fasciculus (part of the language network) were tracked, bilaterally, with constrained spherical deconvolution (CSD). The fan shape of the corticospinal tract, in its upper portion, as well as the C-shape of the arcuate fasciculus, can be recognized.

Sunday, 2 February 2020

fMRI

Functional MRI (fMRI), also called BOLD imaging, is a magnetic resonance imaging-based neuroimaging technique that makes it possible to detect the brain areas that are involved in a task, a process or an emotion.

MRI Brain Perfusion

Perfusion MRI is based on the analysis of the contrast enhancement of MRI images after a peripheral injection of a contrast agent, e.g. Gd-DTPA. The injection of 0.1 mmol/kg is usually quick (i.e. a bolus), at a rate of 5-10 ml/s, and is followed by a saline flush.

Dynamic Susceptibility Contrast (DSC) perfusion MRI uses a GRE-EPI (T2*-weighted) sequence in which the contrast agent induces an hypointensity. For example, 40 volumes can be acquired every 1.5 seconds during 1 minute. Concentration-versus-time curves are calculated from the variation of the MR image signal induced by the contrast agent. The tissue curves are sometimes deconvolved by the arterial input function (AIF) in order to eliminate the influence of the injection kinetics. From the curves, several parameters can be calculated: cerebral blood volume (CBV), cerebral blood flow (CBF), mean transit time (MTT), time to peak (TTP), time to peak of the residue function (Tmax), etc.

Dynamic Contrast Enhanced (DCE) perfusion MRI uses a spoiled fast gradient echo (T1-weighted) sequence in which the contrast agent induces an hyperintensity. Conversion from image signal to concentration-versus-time curves requires a calibration procedure, e.g. a measurement of pre-injection T1 relaxation time by relaxometry. Taking into account the arterial input function (AIF), tracer kinetics models applied to the curves lead to a measurement of the transfer constant between the intra-vascular and extra-vascular spaces Ktrans (a marker of vessel permeability) and of the fractional volume of the extra-vascular extra-cellular space.

DSC perfusion MRI is often used in combination with diffusion MRI for the imaging of ischemic stroke. The mismatch between perfusion and diffusion MRI is a marker of the ischemic penumbra (i.e. the zone which can still be preserved by an appropriate treatment).
In brain tumors, DSC and DCE perfusion can be used to assess the tumoral vascularization, as a marker of the tumor's grade.

Thursday, 30 January 2020

Is it ok for patients to have both an MR and a CT with contrast on the same day


This question was addressed on the RSNA site and there is no issue doing both MRI and CT the same day with Gadolinium and Iodinated Contrast.
Here is more detailed information courtesy of Richard A. Vitti MD from Medical Affairs at GE Healthcare.
CT contrast, or X-ray contrast for IV use, is usually an iodinated compound, whereas most MRI contrast media contains a heavy metal ion usually gadolinium. Iodine provides contrast by virtue of absorption of X-rays at the K-edge. Gd provides contrast in MR by changing the magnetic moment. Generally these agents distribute themselves from the vascular space to the interstitial space of soft tissues, and are elminated by glomerular filtration by the kidneys.

Tuesday, 28 January 2020

TR and TE




TR and TE are basic pulse sequence parameters and stand for repetition time and echo timerespectively. They are typically measured in milliseconds (ms).

Saturday, 25 January 2020

Pulse Sequence Part I

  • Without pulse sequences we can‟t do MRI. Our life depends on it in terms of which kind of image contrast we want to see or, even, which kind of pathology we want to detect.
  • Understanding what a pulse sequence is and how it influences the image is vitally important.
  • A pulse sequence is a sequence of events, which we need to acquire MRI images.
  • These events are: RF pulses, gradient switches and signal collecting.

Pulse Sequence Overview a

Pulse Sequence Overview

  • An MRI sequence is a number of radiofrequency pulses and gradients that result in a set of images with particular appearance.
  • This section presents a simplified approach to recognising and thinking about common MRI sequences, but does not concern itself with the particulars of each sequences.